PG206: Set voltage at HV test is not reached

The set test voltage at high voltage test is not reached.

Set the measuring time to at least 2 seconds, then the voltage will be displayed correctly.

Background: The filter of the measuring display takes one second, which can be a bit too short for a measuring time under 2 seconds. The test voltage itself is always reached, even if less is displayed.